Dionex ICS-5000
+
Installation Instructions
134
Doc. 065447-02 12/14
Calibration Procedure
1. Connect the cell electrical cable to the ED.
2. Connect the reference electrode electrical cable to the ED.
Do not install
the electrode in the cell yet.
3. On the Chromeleon 7 ePanel Set, click the
ED
tab. On the Chromeleon
6.8 panel tabset, click the
EC Detector
tab.
4. Click the
Calibration
button. The Wellness panel opens (see
5. Follow the instructions on the Wellness panel to calibrate the electrode.
6. Disconnect the reference electrode cable and the cell cable.
7. Grasp the cell body and pull straight out to remove the cell from the
detector.
Figure B-8. Wellness Panel: Dionex ICS-5000
+
Electrochemical
Detector (Chromeleon 7 Version Shown)