Dionex ICS-5000
+
Installation Instructions
38
Doc. 065447-02 12/14
5. If you are installing an autosampler, go on to
. If not, go on to
.
3.6
Setting Up the TC
NOTE
The TC cannot be configured for capillary IC applications.
3.6.1
Installing the Temperature Stabilizer in the TC (Optional)
Four column brackets are installed in the TC at the factory. When you finish
this installation procedure, the screws that hold the column brackets in place
will also hold the temperature stabilizers in place.
1. Remove the two screws from each column bracket and remove the
brackets from the TC (see
2. The TC can contain up to two temperature stabilizers (microbore,
P/N 064650; standard bore, P/N 064548). Determine the mounting site
and orientation for each:
•
The temperature stabilizer can be mounted along either the top or
bottom edge of the thermal element (see
•
The mounting holes on the temperature stabilizer are closer to one
side of the stabilizer than the other. Always orient the temperature
stabilizer so that the mounting holes are closer to the
outside edge
of
the thermal element. This ensures that the temperature stabilizer is in
full contact with the thermal element.
Figure 3-9. TC Column Bracket Positions
Column
Bracket
Screws
Thermal
Element