6. Connect the outlet line from the pump to the Dionex EGC
INLET
port.
7. Connect the line from the
ELUENT IN
port of the Dionex RFIC
+
Eluent
Degasser to the Dionex EGC
OUTLET
port (see
temporary connection.
8. Turn over the Dionex EGC (fittings facing down). Shake the cartridge
vigorously and tap it with the palm of your hand 10 to 15 times to
dislodge any air bubbles trapped in the eluent generation chamber.
NOTE
Be sure to dislodge air bubbles each time the Dionex
EGC has been turned upside down.
9. Slide the Dionex EGC into the cartridge holder in the EG.
10. Remove the 10-32 plug from the vent opening on the side of the standard-
pressure Dionex EGC or remove the fitting from the top of the high-
pressure Dionex EGC. Install a 10-32 luer adapter (P/N 063517) in the
opening. The luer adapter is shipped with the cartridge. Remove the luer
cap.
NOTE
Store the 10-32 plug or fitting in a safe place. The plug
or fitting is required when shipping or storing the car-
tridge.