Items Needed for Installing a Conventional Working Electrode
•
ED cell gasket for conventional working electrodes (P/N 045972)
•
Blunt-end tweezers
NOTE
Install the working electrode before connecting the cell to the
ED.
1. Use tweezers to install the gasket over the alignment pins on the cell body
). When correctly installed, one end of the gasket extends
beyond the cell body, to facilitate gasket installation and removal.
2. Verify that the gasket is flat against the cell body and is not wrinkled.
Figure B-5. ED
Cell Gasket for Conventional Working Electrodes
Alignment
Pins
Cell Gasket
for Conventional