1 • Introduction
Doc. 065447-02 12/14
9
illustrates a typical dual-analysis IC system.
1.6
Fitting and Tube Connection Guidelines
Follow these guidelines when connecting tube fittings in the Dionex ICS-
5000
+
system.
•
Use high-pressure fittings, 10-32 fitting bolts (P/N 074449), and high-
pressure 10-32 double-cone ferrules (P/N 074373). These fittings are used
even in low-pressure systems.
•
When connecting tubing to a port, make sure the fitting bolt and ferrule
are at least 2 mm (0.1 in) from the end of the tubing before you insert the
tubing into the port. Do not position the ferrule and fitting bolt flush with
the end of the tubing.
•
If you are installing a capillary system, do not shorten or lengthen the
0.062 mm (0.0025 in) ID blue PEEK
™
(polyether ether ketone) tubing
(capillary tubing) used for Dionex IC Cube connections, and do not
Figure 1-1. Dionex ICS-5000
+
System for Dual-Analysis RFIC-EG
1
2
3
4
Dionex ICS-5000
+
Detector/
Chromatography Module
Dionex ICS-5000
+
Dual
Pump (DP)
Dionex ICS-5000
+
Eluent Generator (EG)
Dionex ICS Series
AS-AP Autosampler
1
2
3
4
Capillary
IC