Dionex ICS-5000
+
Installation Instructions
124
Doc. 065447-02 12/14
8. Remove the black cap from the quick-disconnect fitting, twist off the end
of the quick-disconnect fitting, and then connect the free end of the blue
tubing to a quick-disconnect fitting on the side of the EO Gas Regulator.
NOTE
Low-pressure valves can be installed on the EO regulator
support bracket (see
A.1.3
Pressurizing the Eluent Reservoir
NOTE
Do not pressurize the system until after the reservoir is
plumbed.
1. Turn the gas regulator knob fully counterclockwise to ensure that there is
no pressure to the reservoir when the gas source is turned on.
2. Close the gas shutoff valve on each reservoir cap.
3. Turn on the gas source; the upper limit is 0.34 MPa (50 psi).
Figure A-3. Gas Connections to Reservoir
Gas Source
Tubing Assembly
(P/N 063942)
Quick-
Disconnect
Fitting
One-Way
Valve
Fitting
(Shown in
off
position)