B • Installing DC Options
Doc. 065447-02 12/14
135
8. Install the pH-Ag/AgCl reference electrode in the cell:
•
For a capillary IC system, follow the instructions in the next section.
•
For an analytical IC system, go on to
Installing the pH-Ag/AgCl Reference Electrode in the Cell for Capillary IC
Systems
1. To avoid any hydraulic pressure buildup when inserting the reference
electrode into the cell, make sure that fitting plugs are not installed on the
cell inlet and outlet fittings.
2. Use a sharp tool (for example, the tip of a safety pin or a straight pin) to
remove the pH-Ag/AgCl reference electrode O-ring (see
Figure B-9. pH-Ag/AgCl Reference Electrode O-Ring
Capillary
IC
Reference
Electrode
O-Ring
(Remove for
capillary IC)