A • Installing Pump Options
Doc. 065447-02 12/14
123
50 psi). If needed for your gas source, use the adapter to connect the
tubing to the gas source.
6. For each reservoir to be connected, locate a gas source tubing assembly
(P/N 063942). Each assembly consists of a length of blue, 3 mm (1/8 in)
ID air line tubing with a quick-disconnect fitting on one end and a one-
way valve fitting on the other end (see
7. Push the one-way valve fitting onto the luer fitting on the top of the
reservoir cap. This is the gas shutoff valve for the reservoir.
Figure A-2. EO
Gas Regulator Connections
EO Gas Regulator
(P/N 074425)
Orange Air Line Tubing
(P/N 052297)
GAS IN Fitting
EO Regulator Support
Bracket (P/N 074226)
DC Rear Panel
1
3
2
4
1
2
3
4
5
5