Installation Instructions
148
Doc. 065447-02 12/14
B.4.1
Installing a High-Pressure Valve on the AM
1. While facing the rear of the AM, squeeze the release tabs on the outside
edge of the plug that is installed in the valve mounting hole. Remove the
plug.
2. Unscrew the black locking ring on the front of the high-pressure valve.
3. Attach the appropriate label (6-port or 10-port) to the front of the AM.
4. Remove the two Phillips screws and
washers from the valve.
5. Orient the valve with the model label
6. Hold the valve behind the AM and
insert it through the valve mounting
hole.
7. Reinstall the two Phillips screws and
washers and tighten the screws.
Figure B-25. High-Pressure Valve Installation
8. Replace the black locking ring and tighten it fingertight.
9. Two high-pressure valve cables are connected to the AM prep board (see
). Uncoil the appropriate valve cable and plug it into the
connector on the valve.
Figure B-26. Connections to the AM Prep Board
High-Pressure Valve
Connector (Left)
Reaction Coil Heater
Connector
Low-Pressure Valve
Connector (Left)
High-Pressure Valve
Connector (Right)
Low-Pressure Valve
Connector (Right)
1
2
3
4
5
4
1
2
3
5