1 • Introduction
Doc. 065342-06 12/11
5
Analytical IC pumps operate at flow rates ranging from 0.00 to 10.0 mL/min
and operating pressures up to 35 MPa (5000 psi) without RFIC and up to
21 MPa (3000 psi) with RFIC. Both microbore and standard bore IC
applications are supported.
Dionex ICS-5000 Eluent Generator (EG)
The EG generates high purity acid or base eluents online from deionized
water. The EG can be configured for single- or dual-channel operation. Each
channel includes:
•
A high precision programmable current source (power supply)
For each channel, the following options must be ordered separately for
installation inside the EG.
•
A disposable eluent generator cartridge (EGC). Each cartridge contains an
electrolyte concentrate solution, appropriate for the eluent being
generated.
•
A Continuously Regenerated Trap Column (CR-TC) to remove any
extraneous contaminants from the deionized water source. The CR-TC is
electrolytically-regenerated, which allows it to operate for extended
periods without chemical regeneration.
•
A high-pressure gas removal device that removes electrolysis gases
created during eluent generation
Two versions of EGC and CR-TC are available: one for capillary IC systems
and one for analytical IC systems.
IMPORTANT
If your system includes a capillary pump linked to an
EGC, check the label on the EG degas (capillary)
cartridge in the IC Cube and verify that 5000 psi is
printed on the label (see the figure on the right).
If 5000 psi is not on the label, you must change the high
pressure limit to
21 MPa (3000 psi)
(see
replace the cartridge with an EG degas (capillary)
cartridge that is rated for 35 MPa (5000 psi)
(P/N 074459).
Capillary
IC
5000 psi
P/N 074459
Summary of Contents for Dionex ICS-5000
Page 66: ...Dionex ICS 5000 Operator s Manual 50 Doc 065342 06 12 11 ...
Page 130: ...Dionex ICS 5000 Operator s Manual 114 Doc 065342 06 12 11 ...
Page 168: ...Dionex ICS 5000 Operator s Manual 152 Doc 065342 06 12 11 ...
Page 176: ...Dionex ICS 5000 Operator s Manual 160 Doc 065342 06 12 11 ...
Page 184: ...Dionex ICS 5000 Operator s Manual 168 Doc 065342 06 12 11 ...
Page 200: ...Dionex ICS 5000 Operator s Manual 184 Doc 065342 06 12 11 ...
Page 206: ...Dionex ICS 5000 Operator s Manual 190 Doc 065342 06 12 11 ...
Page 210: ...Dionex ICS 5000 Operator s Manual 194 Doc 065342 06 12 11 ...
Page 214: ...Dionex ICS 5000 Operator s Manual 198 Doc 065342 06 12 11 ...
Page 256: ...Dionex ICS 5000 Operator s Manual 240 Doc 065342 06 12 11 ...
Page 274: ...Dionex ICS 5000 Operator s Manual 258 Doc 065342 06 12 11 ...
Page 318: ...Dionex ICS 5000 Operator s Manual 302 Doc 065342 06 12 11 ...
Page 348: ...Dionex ICS 5000 Operator s Manual 332 Doc 065342 06 12 11 ...
Page 374: ...Dionex ICS 5000 Operator s Manual 358 Doc 065342 06 12 11 ...
Page 422: ...Dionex ICS 5000 Operator s Manual 406 Doc 065342 06 12 11 ...
Page 434: ...Dionex ICS 5000 Operator s Manual 418 Doc 065342 06 12 11 ...
Page 438: ...Dionex ICS 5000 Operator s Manual 422 Doc 065342 06 12 11 ...
Page 448: ...Dionex ICS 5000 Operator s Manual 432 Doc 065342 06 12 11 ...
Page 454: ...Dionex ICS 5000 Operator s Manual 438 Doc 065342 06 12 11 ...
Page 462: ...Dionex ICS 5000 Operator s Manual 446 Doc 065342 06 12 11 ...
Page 482: ...Dionex ICS 5000 Operator s Manual Index 20 Doc 065342 06 12 11 ...