Dionex ICS-5000 Operator’s Manual
158
Doc. 065342-06 12/11
5. If the preferred flow rate is already selected, but the pump flow is off,
click the switch to start the pump at the selected flow rate.
NOTE
After starting the pump or changing the flow rate, wait
at least 5 minutes (longer for flow rates below
1.0 mL/min) before beginning an analysis. This allows
the DP/SP to stabilize the flow rate.
6. Enter the minimum and maximum pressure limits. Setting pressure limits
ensures that the DP/SP will automatically stop if a system malfunction
occurs.
lists the default pressure limits for capillary and
analytical pumps.
Pump Configuration
Default Low
Pressure Limit
Default High
Pressure Limit
Capillary pump not linked
to an EGC
0
41 MPa (6000 psi)
a
a. Capillary IC pumps shipped before February 17, 2011 were rated for lower
operating pressures than the pumps currently being shipped. Refer to the im-
portant note on the next page for details.
Capillary pump linked to an
EGC
1.4 MPa (200 psi)
35 MPa (5000 psi)
b
b. If the installed version of EG degas (capillary) cartridge is not rated for
35 MPa (5000 psi), change the default high pressure limit to 21 MPa
(3000 psi). Refer to the important note on the next page for details.
Analytical pump not linked
to an EGC
0
35 MPa (5000 psi)
Analytical pump linked to
an EGC
1.4 MPa (200 psi)
21 MPa (3000 psi)
Table 4-1. Default Pressure Limits for DP/SP Pumps
Summary of Contents for Dionex ICS-5000
Page 66: ...Dionex ICS 5000 Operator s Manual 50 Doc 065342 06 12 11 ...
Page 130: ...Dionex ICS 5000 Operator s Manual 114 Doc 065342 06 12 11 ...
Page 168: ...Dionex ICS 5000 Operator s Manual 152 Doc 065342 06 12 11 ...
Page 176: ...Dionex ICS 5000 Operator s Manual 160 Doc 065342 06 12 11 ...
Page 184: ...Dionex ICS 5000 Operator s Manual 168 Doc 065342 06 12 11 ...
Page 200: ...Dionex ICS 5000 Operator s Manual 184 Doc 065342 06 12 11 ...
Page 206: ...Dionex ICS 5000 Operator s Manual 190 Doc 065342 06 12 11 ...
Page 210: ...Dionex ICS 5000 Operator s Manual 194 Doc 065342 06 12 11 ...
Page 214: ...Dionex ICS 5000 Operator s Manual 198 Doc 065342 06 12 11 ...
Page 256: ...Dionex ICS 5000 Operator s Manual 240 Doc 065342 06 12 11 ...
Page 274: ...Dionex ICS 5000 Operator s Manual 258 Doc 065342 06 12 11 ...
Page 318: ...Dionex ICS 5000 Operator s Manual 302 Doc 065342 06 12 11 ...
Page 348: ...Dionex ICS 5000 Operator s Manual 332 Doc 065342 06 12 11 ...
Page 374: ...Dionex ICS 5000 Operator s Manual 358 Doc 065342 06 12 11 ...
Page 422: ...Dionex ICS 5000 Operator s Manual 406 Doc 065342 06 12 11 ...
Page 434: ...Dionex ICS 5000 Operator s Manual 418 Doc 065342 06 12 11 ...
Page 438: ...Dionex ICS 5000 Operator s Manual 422 Doc 065342 06 12 11 ...
Page 448: ...Dionex ICS 5000 Operator s Manual 432 Doc 065342 06 12 11 ...
Page 454: ...Dionex ICS 5000 Operator s Manual 438 Doc 065342 06 12 11 ...
Page 462: ...Dionex ICS 5000 Operator s Manual 446 Doc 065342 06 12 11 ...
Page 482: ...Dionex ICS 5000 Operator s Manual Index 20 Doc 065342 06 12 11 ...