Dionex ICS-5000 Operator’s Manual
196
Doc. 065342-06 12/11
NOTE
It is especially important to regularly replace end-line
filters when using aqueous eluents. Aqueous eluents
may contaminate the filters with bacteria or algae,
causing cause flow restrictions to the pump.
End-line filters are provided in the pump ship kit.
Before installation, thoroughly rinse the end-line filter with ASTM Type I
(18 megohm-cm) filtered and deionized water that meets the
specifications listed in
to remove any loose particles.
•
Replace the seal wash solution. ASTM Type I (18 megohm-cm) filtered
and deionized water that meets the specifications listed in
is
appropriate for most applications.
•
Check the tubing connected to the peristaltic pump. If the tubing is
blocked or crimped, replace it (see
•
Check the liquid level in each eluent reservoir and refill as required.
7.4
DP/SP Periodic Maintenance
•
Make fresh eluent as needed.
•
Replace piston seals every 6 to 12 months (see
•
Clean eluent reservoirs every 3 months with methanol or isopropyl
alcohol, as instructed in
7.5
DP/SP Annual Maintenance
Thermo Fisher Scientific recommends performing preventive maintenance
annually. The following kits are available:
•
ICS-5000 Analytical Pump Preventive Maintenance Kit (P/N 075038)
•
ICS-5000 Capillary Pump Preventive Maintenance Kit (P/N 075039)
Pump Ship Kit
Part Number
DP Ship Kit Dual Capillary
072112
DP Ship Kit Dual Analytical
062463
DP Ship Kit Capillary and Analytical
072111
SP Ship Kit Capillary or Analytical
063342
Capillary
IC
Capillary
IC
Summary of Contents for Dionex ICS-5000
Page 66: ...Dionex ICS 5000 Operator s Manual 50 Doc 065342 06 12 11 ...
Page 130: ...Dionex ICS 5000 Operator s Manual 114 Doc 065342 06 12 11 ...
Page 168: ...Dionex ICS 5000 Operator s Manual 152 Doc 065342 06 12 11 ...
Page 176: ...Dionex ICS 5000 Operator s Manual 160 Doc 065342 06 12 11 ...
Page 184: ...Dionex ICS 5000 Operator s Manual 168 Doc 065342 06 12 11 ...
Page 200: ...Dionex ICS 5000 Operator s Manual 184 Doc 065342 06 12 11 ...
Page 206: ...Dionex ICS 5000 Operator s Manual 190 Doc 065342 06 12 11 ...
Page 210: ...Dionex ICS 5000 Operator s Manual 194 Doc 065342 06 12 11 ...
Page 214: ...Dionex ICS 5000 Operator s Manual 198 Doc 065342 06 12 11 ...
Page 256: ...Dionex ICS 5000 Operator s Manual 240 Doc 065342 06 12 11 ...
Page 274: ...Dionex ICS 5000 Operator s Manual 258 Doc 065342 06 12 11 ...
Page 318: ...Dionex ICS 5000 Operator s Manual 302 Doc 065342 06 12 11 ...
Page 348: ...Dionex ICS 5000 Operator s Manual 332 Doc 065342 06 12 11 ...
Page 374: ...Dionex ICS 5000 Operator s Manual 358 Doc 065342 06 12 11 ...
Page 422: ...Dionex ICS 5000 Operator s Manual 406 Doc 065342 06 12 11 ...
Page 434: ...Dionex ICS 5000 Operator s Manual 418 Doc 065342 06 12 11 ...
Page 438: ...Dionex ICS 5000 Operator s Manual 422 Doc 065342 06 12 11 ...
Page 448: ...Dionex ICS 5000 Operator s Manual 432 Doc 065342 06 12 11 ...
Page 454: ...Dionex ICS 5000 Operator s Manual 438 Doc 065342 06 12 11 ...
Page 462: ...Dionex ICS 5000 Operator s Manual 446 Doc 065342 06 12 11 ...
Page 482: ...Dionex ICS 5000 Operator s Manual Index 20 Doc 065342 06 12 11 ...