Dionex ICS-5000 Operator’s Manual
86
Doc. 065342-06 12/11
Recorder
calibration
Zero, Full Scale,
Normal
Use this setting to calibrate a
recording device. Select
Zero
to
set the output signal to zero volts.
Select
Full Scale
to set the
output signal to the selected full-
scale voltage (0.01, 0.10, or
1.00 V). For normal operation,
select
Normal
(the default) to
output a signal corresponding to
the detector output.
Offset level
0 to 100%
Use this setting to adjust the zero
position of the analog output
when it is plotted. The value
entered is a percentage of the
full-scale analog output. An offset
allows a recording device to plot
the signal if it becomes negative.
The offset level does not affect
the magnitude of the output
signal.
Polarity
Positive, Negative
Use this setting to set the polarity
of the analog output signal to
either positive (the default) or
negative. In applications in which
the analyte output is lower than
the background signal, the
polarity must be negative to
display peaks instead of dips on
the chromatogram.
Mark
10% of the full-scale
analog output
Use this setting to send a positive
pulse to the analog output as an
event marker. A mark is typically
used to indicate a sample
injection.
Analog Output
Setting
Values
Description
Table 2-4. Analog Output Configuration Settings (Continued)
Summary of Contents for Dionex ICS-5000
Page 66: ...Dionex ICS 5000 Operator s Manual 50 Doc 065342 06 12 11 ...
Page 130: ...Dionex ICS 5000 Operator s Manual 114 Doc 065342 06 12 11 ...
Page 168: ...Dionex ICS 5000 Operator s Manual 152 Doc 065342 06 12 11 ...
Page 176: ...Dionex ICS 5000 Operator s Manual 160 Doc 065342 06 12 11 ...
Page 184: ...Dionex ICS 5000 Operator s Manual 168 Doc 065342 06 12 11 ...
Page 200: ...Dionex ICS 5000 Operator s Manual 184 Doc 065342 06 12 11 ...
Page 206: ...Dionex ICS 5000 Operator s Manual 190 Doc 065342 06 12 11 ...
Page 210: ...Dionex ICS 5000 Operator s Manual 194 Doc 065342 06 12 11 ...
Page 214: ...Dionex ICS 5000 Operator s Manual 198 Doc 065342 06 12 11 ...
Page 256: ...Dionex ICS 5000 Operator s Manual 240 Doc 065342 06 12 11 ...
Page 274: ...Dionex ICS 5000 Operator s Manual 258 Doc 065342 06 12 11 ...
Page 318: ...Dionex ICS 5000 Operator s Manual 302 Doc 065342 06 12 11 ...
Page 348: ...Dionex ICS 5000 Operator s Manual 332 Doc 065342 06 12 11 ...
Page 374: ...Dionex ICS 5000 Operator s Manual 358 Doc 065342 06 12 11 ...
Page 422: ...Dionex ICS 5000 Operator s Manual 406 Doc 065342 06 12 11 ...
Page 434: ...Dionex ICS 5000 Operator s Manual 418 Doc 065342 06 12 11 ...
Page 438: ...Dionex ICS 5000 Operator s Manual 422 Doc 065342 06 12 11 ...
Page 448: ...Dionex ICS 5000 Operator s Manual 432 Doc 065342 06 12 11 ...
Page 454: ...Dionex ICS 5000 Operator s Manual 438 Doc 065342 06 12 11 ...
Page 462: ...Dionex ICS 5000 Operator s Manual 446 Doc 065342 06 12 11 ...
Page 482: ...Dionex ICS 5000 Operator s Manual Index 20 Doc 065342 06 12 11 ...