2 • EG Description
Doc. 065342-06 12/11
43
RFIC Eluent Degasser
For analytical IC systems, an RFIC Eluent Degasser (P/N 062137) is installed
on the EG component mounting panel. The degasser contains a tubing
assembly that purges the electrolysis gas from the freshly-generated eluent
before it is directed to the separator column.
For capillary IC systems, the degasser is installed in the IC Cube (see
Backpressure Coil (Optional)
For analytical IC systems, the EGC requires at least 14 MPa (2000 psi) of
system backpressure for removal of electrolysis gas from the eluent produced
by the cartridge. A system backpressure of 16 MPa (2300 psi) is ideal.
If necessary, increase the system backpressure by installing a backpressure
coil between the injection valve and the EGC
OUTLET
port. For details, refer
to
.
EPM III Electrolytic pH Modifier and Carbonate Mixer
The EGC III K
2
CO
3
can be used with an EPM III Electrolytic pH Modifier
(P/N 080135) and a carbonate mixer (P/N 061686, 4 mm; P/N 063443, 2 mm)
to produce a carbonate/bicarbonate mixture for use with anion exchange
separations on carbonate-based Dionex IonPac columns.
After the cartridge generates potassium carbonate eluent, the EPM adjusts the
eluent concentration to produce the carbonate/bicarbonate mixture. The
carbonate mixer provides mixing necessary to produce a homogenous
solution of electrolytically-generated K
2
CO
3
and KHCO
3
eluent.
For more information about these products, refer to the EGC manual. EGC
manuals are included on the Thermo Scientific Reference Library DVD
(P/N 053891).
Leak Sensor
If liquid collects in the drip tray in the bottom of the EG, a leak sensor reports
the leak to Chromeleon, and an error message is displayed in the Audit Trail.
In addition, the
Alarm
LED on the EG front panel lights.
Capillary
IC
Summary of Contents for Dionex ICS-5000
Page 66: ...Dionex ICS 5000 Operator s Manual 50 Doc 065342 06 12 11 ...
Page 130: ...Dionex ICS 5000 Operator s Manual 114 Doc 065342 06 12 11 ...
Page 168: ...Dionex ICS 5000 Operator s Manual 152 Doc 065342 06 12 11 ...
Page 176: ...Dionex ICS 5000 Operator s Manual 160 Doc 065342 06 12 11 ...
Page 184: ...Dionex ICS 5000 Operator s Manual 168 Doc 065342 06 12 11 ...
Page 200: ...Dionex ICS 5000 Operator s Manual 184 Doc 065342 06 12 11 ...
Page 206: ...Dionex ICS 5000 Operator s Manual 190 Doc 065342 06 12 11 ...
Page 210: ...Dionex ICS 5000 Operator s Manual 194 Doc 065342 06 12 11 ...
Page 214: ...Dionex ICS 5000 Operator s Manual 198 Doc 065342 06 12 11 ...
Page 256: ...Dionex ICS 5000 Operator s Manual 240 Doc 065342 06 12 11 ...
Page 274: ...Dionex ICS 5000 Operator s Manual 258 Doc 065342 06 12 11 ...
Page 318: ...Dionex ICS 5000 Operator s Manual 302 Doc 065342 06 12 11 ...
Page 348: ...Dionex ICS 5000 Operator s Manual 332 Doc 065342 06 12 11 ...
Page 374: ...Dionex ICS 5000 Operator s Manual 358 Doc 065342 06 12 11 ...
Page 422: ...Dionex ICS 5000 Operator s Manual 406 Doc 065342 06 12 11 ...
Page 434: ...Dionex ICS 5000 Operator s Manual 418 Doc 065342 06 12 11 ...
Page 438: ...Dionex ICS 5000 Operator s Manual 422 Doc 065342 06 12 11 ...
Page 448: ...Dionex ICS 5000 Operator s Manual 432 Doc 065342 06 12 11 ...
Page 454: ...Dionex ICS 5000 Operator s Manual 438 Doc 065342 06 12 11 ...
Page 462: ...Dionex ICS 5000 Operator s Manual 446 Doc 065342 06 12 11 ...
Page 482: ...Dionex ICS 5000 Operator s Manual Index 20 Doc 065342 06 12 11 ...