9 • DC Service
Doc. 065342-06 12/11
389
10. Reinstall the working electrode block and the yoke knob and block.
Fingertighten the yoke knob by turning it a full 360 degrees.
NOTE
It is not possible to overtighten the yoke knob. Once
the knob clicks into place, it does not tighten any fur-
ther. The yoke knob ensures constant pressure is
applied to the cell.
11. Orient the cell assembly with the yoke knob on the left and push the cell
onto its mounting location in the DC compartment.
12. Reconnect the cell’s inlet and outlet lines.
13. Reconnect the electrical cables.
14. Start the pump flow.
15. Wait until the pump pressure has stabilized (30 to 60 seconds) and then
turn on the cell voltage.
9.28.4 Polishing an ED Cell Conventional Working Electrode
These instructions are for conventional (nondisposable) working electrodes
only.
Do not polish disposable electrodes
.
When to Polish the Working Electrode
•
Do not polish new conventional working electrodes before
installation.
•
After an electrode has been used for a period of time, a layer of
contamination may build up. When this occurs, the electrode must be
polished.
•
After the working electrode is polished and installed, background signal
and analyte sensitivity require several hours to stabilize. Once stabilized,
do not polish the electrode unless you observe a loss of signal or severe
electrode recession.
Items Needed
•
Polishing kit (P/N 036313) shipped with the electrode. The kit contains
polishing pads (P/N 036321), a bottle of fine polishing compound
(P/N 036318), and a bottle of coarse polishing compound (P/N 036319).
Summary of Contents for Dionex ICS-5000
Page 66: ...Dionex ICS 5000 Operator s Manual 50 Doc 065342 06 12 11 ...
Page 130: ...Dionex ICS 5000 Operator s Manual 114 Doc 065342 06 12 11 ...
Page 168: ...Dionex ICS 5000 Operator s Manual 152 Doc 065342 06 12 11 ...
Page 176: ...Dionex ICS 5000 Operator s Manual 160 Doc 065342 06 12 11 ...
Page 184: ...Dionex ICS 5000 Operator s Manual 168 Doc 065342 06 12 11 ...
Page 200: ...Dionex ICS 5000 Operator s Manual 184 Doc 065342 06 12 11 ...
Page 206: ...Dionex ICS 5000 Operator s Manual 190 Doc 065342 06 12 11 ...
Page 210: ...Dionex ICS 5000 Operator s Manual 194 Doc 065342 06 12 11 ...
Page 214: ...Dionex ICS 5000 Operator s Manual 198 Doc 065342 06 12 11 ...
Page 256: ...Dionex ICS 5000 Operator s Manual 240 Doc 065342 06 12 11 ...
Page 274: ...Dionex ICS 5000 Operator s Manual 258 Doc 065342 06 12 11 ...
Page 318: ...Dionex ICS 5000 Operator s Manual 302 Doc 065342 06 12 11 ...
Page 348: ...Dionex ICS 5000 Operator s Manual 332 Doc 065342 06 12 11 ...
Page 374: ...Dionex ICS 5000 Operator s Manual 358 Doc 065342 06 12 11 ...
Page 422: ...Dionex ICS 5000 Operator s Manual 406 Doc 065342 06 12 11 ...
Page 434: ...Dionex ICS 5000 Operator s Manual 418 Doc 065342 06 12 11 ...
Page 438: ...Dionex ICS 5000 Operator s Manual 422 Doc 065342 06 12 11 ...
Page 448: ...Dionex ICS 5000 Operator s Manual 432 Doc 065342 06 12 11 ...
Page 454: ...Dionex ICS 5000 Operator s Manual 438 Doc 065342 06 12 11 ...
Page 462: ...Dionex ICS 5000 Operator s Manual 446 Doc 065342 06 12 11 ...
Page 482: ...Dionex ICS 5000 Operator s Manual Index 20 Doc 065342 06 12 11 ...