8 • Troubleshooting
Doc. 065342-06 12/11
215
5. Prepare fresh eluent. To ensure eluent purity, prepare all eluents with
spectro-grade eluents, reagent-grade chemicals, and ASTM Type I
(18 megohm-cm) filtered and deionized water that meets the
specifications listed in
•
DP/SP not properly primed
•
Piston seal is damaged
If the piston seal is damaged, it allows liquid leaks. Leaks are usually
visible, and thus easily detected. If necessary, replace the piston seal (see
•
Inadequate system or cell backpressure
Add backpressure tubing to the cell outlet (see
pump (if no column is installed).
•
(Analytical EG only) System backpressure is below 14 MPa (2000 psi)
The optimal system backpressure is 16 MPa (2300 psi) in an analytical IC
system with an EG installed,. Low system backpressure may cause high
baseline noise as the eluent concentration increases in the gradient. To
correct this, install a backpressure coil in the EG after the RFIC Eluent
Degasser (see
•
Flow system leak ahead of cell
Check all fittings and liquid lines for leaks. Tighten or, if necessary,
replace all liquid line connections (see
•
Rapid changes in ambient temperature
Make sure the column is installed in a thermostatically controlled
compartment and the compartment door is closed.
•
Insufficient system equilibration following changes to operating
parameters; especially apparent when operating at high sensitivities
Allow a longer system equilibration time (up to 2 hours) before starting
operation.
Summary of Contents for Dionex ICS-5000
Page 66: ...Dionex ICS 5000 Operator s Manual 50 Doc 065342 06 12 11 ...
Page 130: ...Dionex ICS 5000 Operator s Manual 114 Doc 065342 06 12 11 ...
Page 168: ...Dionex ICS 5000 Operator s Manual 152 Doc 065342 06 12 11 ...
Page 176: ...Dionex ICS 5000 Operator s Manual 160 Doc 065342 06 12 11 ...
Page 184: ...Dionex ICS 5000 Operator s Manual 168 Doc 065342 06 12 11 ...
Page 200: ...Dionex ICS 5000 Operator s Manual 184 Doc 065342 06 12 11 ...
Page 206: ...Dionex ICS 5000 Operator s Manual 190 Doc 065342 06 12 11 ...
Page 210: ...Dionex ICS 5000 Operator s Manual 194 Doc 065342 06 12 11 ...
Page 214: ...Dionex ICS 5000 Operator s Manual 198 Doc 065342 06 12 11 ...
Page 256: ...Dionex ICS 5000 Operator s Manual 240 Doc 065342 06 12 11 ...
Page 274: ...Dionex ICS 5000 Operator s Manual 258 Doc 065342 06 12 11 ...
Page 318: ...Dionex ICS 5000 Operator s Manual 302 Doc 065342 06 12 11 ...
Page 348: ...Dionex ICS 5000 Operator s Manual 332 Doc 065342 06 12 11 ...
Page 374: ...Dionex ICS 5000 Operator s Manual 358 Doc 065342 06 12 11 ...
Page 422: ...Dionex ICS 5000 Operator s Manual 406 Doc 065342 06 12 11 ...
Page 434: ...Dionex ICS 5000 Operator s Manual 418 Doc 065342 06 12 11 ...
Page 438: ...Dionex ICS 5000 Operator s Manual 422 Doc 065342 06 12 11 ...
Page 448: ...Dionex ICS 5000 Operator s Manual 432 Doc 065342 06 12 11 ...
Page 454: ...Dionex ICS 5000 Operator s Manual 438 Doc 065342 06 12 11 ...
Page 462: ...Dionex ICS 5000 Operator s Manual 446 Doc 065342 06 12 11 ...
Page 482: ...Dionex ICS 5000 Operator s Manual Index 20 Doc 065342 06 12 11 ...