Dionex ICS-5000 Operator’s Manual
308
Doc. 065342-06 12/11
9.4
Cleaning Eluent Reservoirs
Before preparing new eluent, all eluent reservoirs should be rinsed thoroughly
(inside and out) with ASTM Type I (18 megohm-cm) filtered and deionized
water that meets the specifications listed in
. If a reservoir still
appears dirty, or has a slimy film on the inside, clean it as instructed below.
Clean reservoirs in a capillary IC system every 3 months, as instructed below.
1. Dispose of any remaining chemicals according to municipal regulations.
2. Rinse the reservoir (inside and out) with ASTM Type I (18 megohm-cm)
filtered and deionized water.
3. Rinse the inside of the reservoir with isopropyl alcohol or methanol.
4. If algae or bacteria have left a slimy film on the reservoir, use an algicide
or disinfectant (dilute hydrogen peroxide, etc.).
5. Rinse cleaning chemicals out of the reservoir with ASTM Type I
(18 megohm-cm) filtered and deionized water.
6. Dry the reservoir with clean, particulate-free air.
9.5
Priming the DP/SP
The pump should be primed if the eluent has been changed, the eluent line is
new (empty), or the eluent line contains air. This section describes two
methods for priming the pump:
•
To prime the pump with the
PUMP PRIME
•
To prime the pump from the
Pump
panel in Chromeleon, see
.
Although a 10 cc syringe (P/N 054578) can be used with either priming
procedure, using a syringe only if eluent lines are 100% empty or if the pump
is dry is recommended.
Capillary
IC
Summary of Contents for Dionex ICS-5000
Page 66: ...Dionex ICS 5000 Operator s Manual 50 Doc 065342 06 12 11 ...
Page 130: ...Dionex ICS 5000 Operator s Manual 114 Doc 065342 06 12 11 ...
Page 168: ...Dionex ICS 5000 Operator s Manual 152 Doc 065342 06 12 11 ...
Page 176: ...Dionex ICS 5000 Operator s Manual 160 Doc 065342 06 12 11 ...
Page 184: ...Dionex ICS 5000 Operator s Manual 168 Doc 065342 06 12 11 ...
Page 200: ...Dionex ICS 5000 Operator s Manual 184 Doc 065342 06 12 11 ...
Page 206: ...Dionex ICS 5000 Operator s Manual 190 Doc 065342 06 12 11 ...
Page 210: ...Dionex ICS 5000 Operator s Manual 194 Doc 065342 06 12 11 ...
Page 214: ...Dionex ICS 5000 Operator s Manual 198 Doc 065342 06 12 11 ...
Page 256: ...Dionex ICS 5000 Operator s Manual 240 Doc 065342 06 12 11 ...
Page 274: ...Dionex ICS 5000 Operator s Manual 258 Doc 065342 06 12 11 ...
Page 318: ...Dionex ICS 5000 Operator s Manual 302 Doc 065342 06 12 11 ...
Page 348: ...Dionex ICS 5000 Operator s Manual 332 Doc 065342 06 12 11 ...
Page 374: ...Dionex ICS 5000 Operator s Manual 358 Doc 065342 06 12 11 ...
Page 422: ...Dionex ICS 5000 Operator s Manual 406 Doc 065342 06 12 11 ...
Page 434: ...Dionex ICS 5000 Operator s Manual 418 Doc 065342 06 12 11 ...
Page 438: ...Dionex ICS 5000 Operator s Manual 422 Doc 065342 06 12 11 ...
Page 448: ...Dionex ICS 5000 Operator s Manual 432 Doc 065342 06 12 11 ...
Page 454: ...Dionex ICS 5000 Operator s Manual 438 Doc 065342 06 12 11 ...
Page 462: ...Dionex ICS 5000 Operator s Manual 446 Doc 065342 06 12 11 ...
Page 482: ...Dionex ICS 5000 Operator s Manual Index 20 Doc 065342 06 12 11 ...