The temperature achieved may vary from the control range, depending on the
ambient temperature (see the table below).
For example, if the ambient air temperature in the lab is 30 °C and you set the
temperature to 5 °C, the actual temperature achieved will be about 12 °C.
2.26.1 Predictive Performance
The Predictive Performance feature of the TC provides various functions for
estimating the lifetime of replaceable parts and for planning and recording
service and qualification procedures.
A partial list of the TC Predictive Performance functions is provided below.
Refer to the Chromeleon Help for details.
•
Monitoring of the total cooling and heating workload performed by the
compartment and the total operating hours since initial operation
•
Monitoring of the injection valve usage
•
Reminders for service and qualification periods, including display of
warning messages when limits are approaching or overdue
Predictive Performance commands and parameters are available in the
Chromeleon Commands dialog box.
Control Range
Achieved Temperatures (Based on Ambient)
5 to 85 °C
Minimum temperature 18 °C below ambient
Maximum temperature 70 °C above ambient
Do not touch any metal parts inside the TC while it is heating up or after
it reaches the set point temperature. Wait for the compartment to cool
down before servicing any parts.
Ne touchez à aucune des pièces métalliques à l'intérieur du TC pendant
qu'il chauffe ou après qu'il ait atteint la température de consigne.
Attendez que le compartiment soit refroidi avant de réparer les pièces.
Berühren Sie im TC keine metallischen Teile, während dieses
aufgeheizt wird oder nachdem die Solltemperatur erreicht ist. Warten
Sie, bis die Kammer abgekühlt ist, ehe Sie Wartungsarbeiten
durchführen.
Summary of Contents for Dionex ICS-5000
Page 66: ...Dionex ICS 5000 Operator s Manual 50 Doc 065342 06 12 11 ...
Page 130: ...Dionex ICS 5000 Operator s Manual 114 Doc 065342 06 12 11 ...
Page 168: ...Dionex ICS 5000 Operator s Manual 152 Doc 065342 06 12 11 ...
Page 176: ...Dionex ICS 5000 Operator s Manual 160 Doc 065342 06 12 11 ...
Page 184: ...Dionex ICS 5000 Operator s Manual 168 Doc 065342 06 12 11 ...
Page 200: ...Dionex ICS 5000 Operator s Manual 184 Doc 065342 06 12 11 ...
Page 206: ...Dionex ICS 5000 Operator s Manual 190 Doc 065342 06 12 11 ...
Page 210: ...Dionex ICS 5000 Operator s Manual 194 Doc 065342 06 12 11 ...
Page 214: ...Dionex ICS 5000 Operator s Manual 198 Doc 065342 06 12 11 ...
Page 256: ...Dionex ICS 5000 Operator s Manual 240 Doc 065342 06 12 11 ...
Page 274: ...Dionex ICS 5000 Operator s Manual 258 Doc 065342 06 12 11 ...
Page 318: ...Dionex ICS 5000 Operator s Manual 302 Doc 065342 06 12 11 ...
Page 348: ...Dionex ICS 5000 Operator s Manual 332 Doc 065342 06 12 11 ...
Page 374: ...Dionex ICS 5000 Operator s Manual 358 Doc 065342 06 12 11 ...
Page 422: ...Dionex ICS 5000 Operator s Manual 406 Doc 065342 06 12 11 ...
Page 434: ...Dionex ICS 5000 Operator s Manual 418 Doc 065342 06 12 11 ...
Page 438: ...Dionex ICS 5000 Operator s Manual 422 Doc 065342 06 12 11 ...
Page 448: ...Dionex ICS 5000 Operator s Manual 432 Doc 065342 06 12 11 ...
Page 454: ...Dionex ICS 5000 Operator s Manual 438 Doc 065342 06 12 11 ...
Page 462: ...Dionex ICS 5000 Operator s Manual 446 Doc 065342 06 12 11 ...
Page 482: ...Dionex ICS 5000 Operator s Manual Index 20 Doc 065342 06 12 11 ...