6. Press
PUMP PRIME 1
(or
PUMP PRIME 2
) on the front of the DP/SP. The
pump will begin pumping at the priming flow rate selected on the
Chromeleon panel.
7. Prime the pump until all air is purged and no air bubbles can be seen
exiting the waste line.
8.
Gradient pump only:
To prime additional eluent lines, select another
eluent channel and repeat the priming procedure.
9. After priming all eluent lines, press
PUMP PRIME 1
(or
PUMP PRIME 2
) to
stop priming and return to the flow rate last selected.
10. Close the priming valve by turning it clockwise.
Do not overtighten the
priming valve.
11. Reconnect the pump to the software by clicking
Connected
on the
Pump
panel.
9.5.2
Priming from the Chromeleon Panel
1. Open the priming valve on the secondary pump head (see
turning it one-half turn counterclockwise.
NOTE
If the priming valve is opened too much, air is drawn
through the valve and air bubbles can be seen exiting
the waste line.
2. On the Chromeleon 7 ePanel Set or the Chromeleon 6.8 panel tabset, click
the
Pump
tab.
3.
Gradient pump only:
Enter 100% for the channel (A, B, C, or D) to be
primed.
4. Set the priming flow rate to the desired rate. The default rate for a
capillary IC pump is 3.0 mL/min. The default rate for an analytical IC
pump is 6.0 mL/min.
5. Click the
Prime
button on the
Pump
panel. A warning message asks you
to verify that the purge (priming) valve is open. Click
Execute despite
warnings
(Chromeleon 7) or
OK
(Chromeleon 6.8). Priming starts.
Do not use any tools to tighten the priming valve! Overtightening may
destroy the cap seal. Open or close the priming valve only when the
system pressure is down.
Summary of Contents for Dionex ICS-5000
Page 66: ...Dionex ICS 5000 Operator s Manual 50 Doc 065342 06 12 11 ...
Page 130: ...Dionex ICS 5000 Operator s Manual 114 Doc 065342 06 12 11 ...
Page 168: ...Dionex ICS 5000 Operator s Manual 152 Doc 065342 06 12 11 ...
Page 176: ...Dionex ICS 5000 Operator s Manual 160 Doc 065342 06 12 11 ...
Page 184: ...Dionex ICS 5000 Operator s Manual 168 Doc 065342 06 12 11 ...
Page 200: ...Dionex ICS 5000 Operator s Manual 184 Doc 065342 06 12 11 ...
Page 206: ...Dionex ICS 5000 Operator s Manual 190 Doc 065342 06 12 11 ...
Page 210: ...Dionex ICS 5000 Operator s Manual 194 Doc 065342 06 12 11 ...
Page 214: ...Dionex ICS 5000 Operator s Manual 198 Doc 065342 06 12 11 ...
Page 256: ...Dionex ICS 5000 Operator s Manual 240 Doc 065342 06 12 11 ...
Page 274: ...Dionex ICS 5000 Operator s Manual 258 Doc 065342 06 12 11 ...
Page 318: ...Dionex ICS 5000 Operator s Manual 302 Doc 065342 06 12 11 ...
Page 348: ...Dionex ICS 5000 Operator s Manual 332 Doc 065342 06 12 11 ...
Page 374: ...Dionex ICS 5000 Operator s Manual 358 Doc 065342 06 12 11 ...
Page 422: ...Dionex ICS 5000 Operator s Manual 406 Doc 065342 06 12 11 ...
Page 434: ...Dionex ICS 5000 Operator s Manual 418 Doc 065342 06 12 11 ...
Page 438: ...Dionex ICS 5000 Operator s Manual 422 Doc 065342 06 12 11 ...
Page 448: ...Dionex ICS 5000 Operator s Manual 432 Doc 065342 06 12 11 ...
Page 454: ...Dionex ICS 5000 Operator s Manual 438 Doc 065342 06 12 11 ...
Page 462: ...Dionex ICS 5000 Operator s Manual 446 Doc 065342 06 12 11 ...
Page 482: ...Dionex ICS 5000 Operator s Manual Index 20 Doc 065342 06 12 11 ...