2 • DC Description
Doc. 065342-06 12/11
85
2.17.2 Analog Outputs
When the I/O option is installed, two analog outputs (one for each detector)
are installed on the DC rear panel (see
). The analog outputs
supply a voltage signal proportional to the current measured by the detector
cell. The outputs can be connected to an analog-to-digital (A/D) converter
such as an integrator or other recording device. Refer to
the documentation for the device for connection instructions.
Several settings are available that allow you to configure the analog output
signal for your detector and connected device.
describes the
settings. You select the preferred settings on the detector panel on the
Chromeleon 7 ePanel Set (see
) or Chromeleon 6.8 panel tabset.
Analog Output
Setting
Values
Description
Full-scale voltage
0.01, 0.10, or 1.00 V
Sets the voltage output of a full-
scale detector response. The
voltage to use depends on the
recording device to which the
analog output is connected. For
example, if the analog output is
connected to a device that
accepts input voltages up to 1 V,
select a full-scale voltage output
of 1 V.
Range
Conductivity:
0.01 to 15,000
S
DC Amperometry:
50 pA to 300
A
Integrated
Amperometry:
50 pC to 200
C
Sets the full-scale detector
response value. The range to use
depends on the detector readings
expected for the application. For
example, selecting a range of
20 micro Siemens (
S) will limit
you to viewing conductivity
readings of 20
S or less.
Table 2-4. Analog Output Configuration Settings
Summary of Contents for Dionex ICS-5000
Page 66: ...Dionex ICS 5000 Operator s Manual 50 Doc 065342 06 12 11 ...
Page 130: ...Dionex ICS 5000 Operator s Manual 114 Doc 065342 06 12 11 ...
Page 168: ...Dionex ICS 5000 Operator s Manual 152 Doc 065342 06 12 11 ...
Page 176: ...Dionex ICS 5000 Operator s Manual 160 Doc 065342 06 12 11 ...
Page 184: ...Dionex ICS 5000 Operator s Manual 168 Doc 065342 06 12 11 ...
Page 200: ...Dionex ICS 5000 Operator s Manual 184 Doc 065342 06 12 11 ...
Page 206: ...Dionex ICS 5000 Operator s Manual 190 Doc 065342 06 12 11 ...
Page 210: ...Dionex ICS 5000 Operator s Manual 194 Doc 065342 06 12 11 ...
Page 214: ...Dionex ICS 5000 Operator s Manual 198 Doc 065342 06 12 11 ...
Page 256: ...Dionex ICS 5000 Operator s Manual 240 Doc 065342 06 12 11 ...
Page 274: ...Dionex ICS 5000 Operator s Manual 258 Doc 065342 06 12 11 ...
Page 318: ...Dionex ICS 5000 Operator s Manual 302 Doc 065342 06 12 11 ...
Page 348: ...Dionex ICS 5000 Operator s Manual 332 Doc 065342 06 12 11 ...
Page 374: ...Dionex ICS 5000 Operator s Manual 358 Doc 065342 06 12 11 ...
Page 422: ...Dionex ICS 5000 Operator s Manual 406 Doc 065342 06 12 11 ...
Page 434: ...Dionex ICS 5000 Operator s Manual 418 Doc 065342 06 12 11 ...
Page 438: ...Dionex ICS 5000 Operator s Manual 422 Doc 065342 06 12 11 ...
Page 448: ...Dionex ICS 5000 Operator s Manual 432 Doc 065342 06 12 11 ...
Page 454: ...Dionex ICS 5000 Operator s Manual 438 Doc 065342 06 12 11 ...
Page 462: ...Dionex ICS 5000 Operator s Manual 446 Doc 065342 06 12 11 ...
Page 482: ...Dionex ICS 5000 Operator s Manual Index 20 Doc 065342 06 12 11 ...