Dionex ICS-5000 Operator’s Manual
186
Doc. 065342-06 12/11
2. Verify that the current to the EGC and the suppressor is off.
3. If you plan to store the EGC in the EG, no further action is required.
If you plan to store the cartridge outside the EG (for example, in the
original shipping container), follow the instructions in
, to
remove the EGC from the EG, and the instructions in
, to
store the EGC.
To restart the EG:
1. Select the parameters for the analysis on the Chromeleon 7 ePanel Set or
the Chromeleon 6.8 panel tabset. For EG startup information, see
.
2. Condition the EGC. For instructions, see
“Conditioning the Analytical EGC” on page 348
.
3. Hydrate the CR-TC (see
“Installing and Flushing the New CR-TC” on
4. Hydrate the suppressor as instructed in the product manual. Suppressor
manuals are included on the Thermo Scientific Reference Library DVD
(P/N 053891).
5. Let the system equilibrate before collecting data.
Before shipping the EG:
1. Follow the instructions in
, to remove the EGC from the
EG.
2. Plug all fittings. Tighten the vent fitting (capillary EGC) or cap the vent
port (analytical EGC). This will prevent the evaporation of water in the
electrolyte reservoir.
Supplying power to the EGC when there is no eluent flow can severely
damage the cartridge.
Make sure the EGC packaging conforms to hazardous material
shipping requirements. Refer to the Material Safety Data Sheet (MSDS)
shipped with the EGC for the chemical description.
Summary of Contents for Dionex ICS-5000
Page 66: ...Dionex ICS 5000 Operator s Manual 50 Doc 065342 06 12 11 ...
Page 130: ...Dionex ICS 5000 Operator s Manual 114 Doc 065342 06 12 11 ...
Page 168: ...Dionex ICS 5000 Operator s Manual 152 Doc 065342 06 12 11 ...
Page 176: ...Dionex ICS 5000 Operator s Manual 160 Doc 065342 06 12 11 ...
Page 184: ...Dionex ICS 5000 Operator s Manual 168 Doc 065342 06 12 11 ...
Page 200: ...Dionex ICS 5000 Operator s Manual 184 Doc 065342 06 12 11 ...
Page 206: ...Dionex ICS 5000 Operator s Manual 190 Doc 065342 06 12 11 ...
Page 210: ...Dionex ICS 5000 Operator s Manual 194 Doc 065342 06 12 11 ...
Page 214: ...Dionex ICS 5000 Operator s Manual 198 Doc 065342 06 12 11 ...
Page 256: ...Dionex ICS 5000 Operator s Manual 240 Doc 065342 06 12 11 ...
Page 274: ...Dionex ICS 5000 Operator s Manual 258 Doc 065342 06 12 11 ...
Page 318: ...Dionex ICS 5000 Operator s Manual 302 Doc 065342 06 12 11 ...
Page 348: ...Dionex ICS 5000 Operator s Manual 332 Doc 065342 06 12 11 ...
Page 374: ...Dionex ICS 5000 Operator s Manual 358 Doc 065342 06 12 11 ...
Page 422: ...Dionex ICS 5000 Operator s Manual 406 Doc 065342 06 12 11 ...
Page 434: ...Dionex ICS 5000 Operator s Manual 418 Doc 065342 06 12 11 ...
Page 438: ...Dionex ICS 5000 Operator s Manual 422 Doc 065342 06 12 11 ...
Page 448: ...Dionex ICS 5000 Operator s Manual 432 Doc 065342 06 12 11 ...
Page 454: ...Dionex ICS 5000 Operator s Manual 438 Doc 065342 06 12 11 ...
Page 462: ...Dionex ICS 5000 Operator s Manual 446 Doc 065342 06 12 11 ...
Page 482: ...Dionex ICS 5000 Operator s Manual Index 20 Doc 065342 06 12 11 ...