Dionex ICS-5000 Operator’s Manual
110
Doc. 065342-06 12/11
time (in minutes) is on the x-axis and the detector response (in
nanoCoulombs) is on the y-axis.
For 3D amperometry data, Chromeleon stores raw data points at a rate of
1 KHz throughout the entire run. Each raw 3D data point is defined by the
following attributes: the detector current (1) recorded at a particular
waveform time (2) and at a specific retention time (3). As a result, the two-
dimensional view of an integrated amperometry chromatogram (retention
time vs. integrated current) is extended by a third dimension (waveform time).
Thus, data for the entire waveform period—not just the integration
intervals—are collected.
Displaying and Reprocessing 3D Data
This section describes how to display and reprocess 3D amperometry data
using Chromeleon 6.8. If you are using Chromeleon 7, the features described
here are available by displaying the data in a contour plot. For details, refer to
the Chromeleon 7 Help.
In Chromeleon 6.8 you can display and reprocess 3D amperometry data in the
3D Amperometry View window (see
). To open the window,
Figure 2-56. Example Chromatogram for a Pulsed Amperometry Application
0
1
2
3
4
5
6
Minutes
0
5
10
40
Analytical Column: CarboPac PA10
and
Guard
Eluent:
18 mM NaOH
Flow Rate:
1.5 mL/min
Detection:
Pulsed Amperometry
Electrode:
Gold
Waveform:
Quadruple Potential
(4 x 250 mm)
Peaks:
1. L-Fucose
2. D-Galactosamine
3. D-Glucosamine
4. D-Galactose
5. D-Glucose
6. D-Mannose
Retention Time (Minutes)
Response
(nC)
Summary of Contents for Dionex ICS-5000
Page 66: ...Dionex ICS 5000 Operator s Manual 50 Doc 065342 06 12 11 ...
Page 130: ...Dionex ICS 5000 Operator s Manual 114 Doc 065342 06 12 11 ...
Page 168: ...Dionex ICS 5000 Operator s Manual 152 Doc 065342 06 12 11 ...
Page 176: ...Dionex ICS 5000 Operator s Manual 160 Doc 065342 06 12 11 ...
Page 184: ...Dionex ICS 5000 Operator s Manual 168 Doc 065342 06 12 11 ...
Page 200: ...Dionex ICS 5000 Operator s Manual 184 Doc 065342 06 12 11 ...
Page 206: ...Dionex ICS 5000 Operator s Manual 190 Doc 065342 06 12 11 ...
Page 210: ...Dionex ICS 5000 Operator s Manual 194 Doc 065342 06 12 11 ...
Page 214: ...Dionex ICS 5000 Operator s Manual 198 Doc 065342 06 12 11 ...
Page 256: ...Dionex ICS 5000 Operator s Manual 240 Doc 065342 06 12 11 ...
Page 274: ...Dionex ICS 5000 Operator s Manual 258 Doc 065342 06 12 11 ...
Page 318: ...Dionex ICS 5000 Operator s Manual 302 Doc 065342 06 12 11 ...
Page 348: ...Dionex ICS 5000 Operator s Manual 332 Doc 065342 06 12 11 ...
Page 374: ...Dionex ICS 5000 Operator s Manual 358 Doc 065342 06 12 11 ...
Page 422: ...Dionex ICS 5000 Operator s Manual 406 Doc 065342 06 12 11 ...
Page 434: ...Dionex ICS 5000 Operator s Manual 418 Doc 065342 06 12 11 ...
Page 438: ...Dionex ICS 5000 Operator s Manual 422 Doc 065342 06 12 11 ...
Page 448: ...Dionex ICS 5000 Operator s Manual 432 Doc 065342 06 12 11 ...
Page 454: ...Dionex ICS 5000 Operator s Manual 438 Doc 065342 06 12 11 ...
Page 462: ...Dionex ICS 5000 Operator s Manual 446 Doc 065342 06 12 11 ...
Page 482: ...Dionex ICS 5000 Operator s Manual Index 20 Doc 065342 06 12 11 ...