The Object Name column lists the items that require signatures:
If the Object Name
lists
The object is
For information, see
A dye set name
The most recent dye calibration
for the dye set specified in the
plate setup.
An install standard
name
The most recent install run for
the application specified in the
plate setup.
A plate name
Plate setup
The example below indicates that E-signature is configured to require 4 signatures
when you start a run, 1 signature is complete and 3 signatures are pending:
• 2 administrator signatures for dye calibration (dye set name is listed two times):
– 1 signature that specifies the "Reviewed and approved" E-signature meaning
(complete, because 1 signature is required [Signatures Required=1] and 1
signature has been logged [Current=1]).
– 1 signature that specifies the "Checked consumables" meaning (incomplete).
• 1 administrator signature for the plate setup that specifies the "Reviewed and
approved" E-signature meaning (incomplete).
• 1 administrator signature for install run that specifies the "Reviewed and
approved" E-signature meaning (incomplete).
Chapter 4
Use the instrument with the Security, Auditing, and E
‑
signature (SAE) v2.0 module
E-signature requirements when you start a run