3.
Touch Detail to see allele-specific details for fragment/HID analysis and basecall
accuracy for sequencing.
4.
Touch Export report (see “Export the install run report“ on page 208).
Fragment analysis install run pass/fail criteria
Result
Acceptance criteria
# Allele peaks
15 for all four capillaries
# Size standard peaks
34 for all four capillaries
Standard deviation of the observed allele fragment sizes
for all 4 capillaries
<0.15 base pairs (bp)
Appendix C
Dye calibration and install standard checks
Perform an install run
C
SeqStudio
™
Genetic Analyzer Instrument and Software User Guide
207