Perform an install run
Install checks are performed by a Field Service Engineer at the time of installation.
An install check can be run at any time with the following reagents to ensure
instrument performance:
Install check type
Reagent
Sequencing
• 3500/3500xL Sequencing Standards, BigDye
™
Terminator v3.1
(Cat. No. 4404312)
• 3500/3500xL Sequencing Standards, BigDye
™
Terminator v1.1
(Cat. No. 4404314)
Fragment
DS
‑
33 GeneScan
™
Installation Standards with GeneScan
™
600 LIZ
™
Size Standard v2.0 (Cat. No. 4376911)
HID
Note:
A J6 manual
dye calibration is
required before the
install check.
GlobalFiler
™
Allelic Ladder from one of the following installation
kits:
• SeqStudio
™
HID and Sequencing Installation Kit (Cat. No.
A46180)
• SeqStudio
™
HID Installation Kit (Cat. No. A46182)
You can include multiple injection groups, multiple applications, and/or multiple
chemistries on an install run plate. For example, you can prepare an install run plate
that contains replictate injections of the fragment install standard and both
sequencing standards.
Figure 22
Example install run plate with multiple applications and replicate injections
Overview of install
checks
Appendix C
Dye calibration and install standard checks
Perform an install run
C
SeqStudio
™
Genetic Analyzer Instrument and Software User Guide
203