2.
Prepare the cartridge.
a.
Remove the cartridge from the box.
Note:
Save the box for off-instrument cartridge storage.
Remove Before
Insertion
b.
Remove the optical cover from the
cartridge by grasping the finger
holds, then pulling toward you.
Note:
Save the optical cover for
off-instrument cartridge storage.
Avoid touching the capillaries that
are protected by the optical cover.
c.
Pinch the clamp on the
SeqStudio
™
Integrated Capillary
Protector, then pull down.
Discard the SeqStudio
™
Integrated
Capillary Protector. The
SeqStudio
™
Integrated Capillary
Protector is single-use. Use a new
SeqStudio
™
Integrated Capillary
Protector if you store the
cartridge.
1
1
Optical cover