Link the instrument to your Connect
cloud-based platform account—
detailed instructions
■
Workflow: Set up the instrument for Connect cloud-based platform access . . 184
■
Network requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184
■
Link the instrument from a mobile device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185
■
Link the instrument using a link code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 186
■
Set up email notifications from the instrument . . . . . . . . . . . . . . . . . . . . . . . . . . 190
Workflow: Set up the instrument for Connect cloud-based platform
access
Register and obtain a Connect cloud-based platform account
▼
Link the instrument to your Connect cloud-based platform
account in any of the following ways:
• Link the instrument from a mobile device
• Link the instrument using a link code
Network requirements
The instrument is factory-configured for IPv4 TCP/IP communication and includes a
fast Ethernet adapter (10/100 Mbps) with a RJ45-type connector for integrating the
device into a local area network (LAN).
By default, the instrument is configured to use the Dynamic Host Configuration
Protocol (DHCP) but can use a static IP address.
The instrument should be configured behind a firewall. Contact Support for
information on required firewall exceptions.
B
184
SeqStudio
™
Genetic Analyzer Instrument and Software User Guide