HID analysis install run pass/fail criteria
Result
Acceptance criteria
# Allele peaks
343 for all four capillaries
# Size standard peaks
26 for all four capillaries
Minimum peak height
>400 RFU
Standard deviation of the observed allele fragment sizes
for all 4 capillaries
<0.15 base pairs (bp)
Sequence analysis install run pass/fail criteria
Result
Acceptance criteria
QV30 CRL (contiguous read
length)
Note:
The contiguous read
length passing criteria for
install checks is an
uninterrupted segment of
bases with an average Quality
Value (QV) of 30.
BigDye
™
Terminator v3.1
Capillaries with QV30 CRL
≥
500 bp pass.
BigDye
™
Terminator v1.1
Capillaries with a QV30 CRL
≥
500 bp pass.
The remaining results on the screen are for information only.
1.
Touch Settings
4
Maintenance and Service
4
Install run
4
Install run history.
2.
Touch the install run of interest.
3.
Touch Export report.
4.
Select a destination for the report.
5.
Touch OK.
Export the install
run report
Appendix C
Dye calibration and install standard checks
Perform an install run
C
208
SeqStudio
™
Genetic Analyzer Instrument and Software User Guide