3
Hardware Upgrades
Dynamic Mixer Installation
46
Accela 600 and 1250 Pump Hardware Manual
Thermo Scientific
Connecting the Dynamic Mixer to the Controller Board
To connect the dynamic mixer to the Controller board
Connect the dynamic mixer’s cable to the P3 cable (see
). One end of the P3
cable is connected to the Controller board and the other end is free inside the pump
chassis.
Figure 34.
Dynamic mixer connection to the P3 cable
Go to the next procedure,
“Reconnecting the Pump Chassis to the Enclosure.”
Reconnecting the Pump Chassis to the Enclosure
To reconnect the pump chassis to the enclosure
1. Push the pump chassis back into the enclosure, and then secure the chassis to the
enclosure with the four screws that you removed in
.
2. Reconnect the LED cable to the LED board.
3. Align the hole in the LED cover plate with the port in the door for the screw. Slide the
screw through the hole. Using a #2 Phillips head screwdriver, tighten the screw to secure
the metal plate to the left door of the pump.
To complete the installation of the dynamic mixer, go to the next procedure,
the Static Mixer and the High-Pressure Lines.”
Dynamic mixer
cable
P3
cable