5
Maintenance
Replacing the Primary Piston Seals
Thermo Scientific
Accela 600 and 1250 Pump Hardware Manual
99
10. Insert the piston into the bore of the displacement chamber. Do not apply any side load
onto the seal during this process (see
Figure 99.
Using the guide bushing and piston to insert the piston seal
11. Hold down the guide bushing and slide the piston up until it is out of the seal
(see
Figure 100.
Sliding the piston out of the seal