1
Introduction
Status LEDs
Thermo Scientific
Accela 600 and 1250 Pump Hardware Manual
7
Status LEDs
The 600 and 1250 pumps have four status LEDs on the left door that provide information
about the pump.
lists the states of the four status LEDs.
Table 3.
Status LEDs for the pump
LED
Status
Description
Power
Green
The pump is switched on and is receiving power.
Comm
Amber
The pump is not communicating with the data system
computer.
Green
Communication to the data system computer has been
established.
Flashing green
A pump program is downloading from the data system
computer.
Run
Amber
The power is switched on, but the pump pistons are idle,
producing no flow.
Flashing amber
A firmware download is in progress or an error condition
has occurred.
Green
The pump pistons are moving, but a pump program is not
running. This can occur when the pump is under direct
control or a pump program has ended.
Flashing green
The pump is running a pump program from a downloaded
method.
Degas
Amber
The degas unit is building vacuum.
Flashing amber
A failure, such as a loss of vacuum, has occurred.
For additional diagnostics information, see
.
Green
Sufficient vacuum has developed to perform liquid
chromatography.