Accela 600 Pump or Accela 1250 Pump Installation
Back Panel Connections
22
Accela 600 and 1250 Pump Hardware Manual
Thermo Scientific
Contact Closure Connections for Accela Systems with an Accela Open Autosampler
The contact closure cable for an Accela LC system with an Accela Open Autosampler is
provided with the Accela Open Autosampler. LC/MS systems that include a TSQ™ Series or
Exactive™ mass spectrometer or an Accela detector require an additional adapter cable, which
is also supplied with the Accela Open Autosampler.
shows the contact closure cable for an Accela LC system with an Accela Open
Autosampler and one or two Accela 600 Pumps, Accela 1250 Pumps, or a combination of
these pumps.
Figure 10.
Accela Open Autosampler interconnect cable (P/N 60157-63024)
shows the adapter cable for an Accela LC system with an Accela Open Autosampler
and an Accela detector (refractive index, UV-Vis, or photodiode array detector) or TSQ Series
or Exactive mass spectrometer.
Tip
An unlabeled cable equivalent to the Accela Open Autosampler interconnect cable
(P/N 60157-63024) ships with the autosampler.
This unlabeled cable has three connectors:
• The DB15 connector plugs into the Interface 1 port on the back panel of the
autosampler.
• The DB25 connector plugs into the D-Sub 25 port on the back panel of an
Accela 600 or 1250 Pump. For a dual-pump system, plug this connector into the
eluting pump’s D-Sub 25 port. Do not connect a contact closure cable to the loading
pump.
• The two-socket connector connects to the Start In pins of an MSQ™ or LTQ™ Series
mass spectrometer.