5
Maintenance
Accessing, Disassembling, and Reassembling the LDA
Thermo Scientific
Accela 600 and 1250 Pump Hardware Manual
83
8. Pull out the two piston retaining springs (see
).
Figure 71.
Piston retaining springs
9. Gently push the piston heads to the left so that they do not contact the Z-arms
(see
Figure 72.
Piston heads pushed to the left, away from the Z-arms
Piston retaining springs
Piston head
Z-arm