5
Maintenance
Accessing, Disassembling, and Reassembling the LDA
Thermo Scientific
Accela 600 and 1250 Pump Hardware Manual
93
Figure 88.
Mounting the LDA onto the pump
4. Using a 4 mm hex ball driver, tighten the three LDA attachment screws (see
).
Figure 89.
LDA attachment screws
Alignment pins
CAUTION
When tightening the screw near the inlet valve, do not touch or apply
force on inlet valve D or it might break.
LDA
attachment
screws
Valve D
(in front of the
lower right
attachment
screw)