2
Accela 600 Pump or Accela 1250 Pump Installation
Solvent Line Connections
32
Accela 600 and 1250 Pump Hardware Manual
Thermo Scientific
shows the two-piece, compression fitting. The length of the nut is 0.45 inches,
and the length of the ferrule is 0.19 inches.
Figure 18.
Two-piece stainless steel fitting
You can find these fittings in the Accela System Kit. You can also order replacement
ferrules from VIC Valco™ Instruments Company and replacement nuts from IDEX™
Corporation.
3. Using a 1/4 in. open-ended wrench, tighten the nut.
Connecting the High-Pressure Solvent Line to the Autosampler
To connect the high-pressure solvent line, follow the procedure for your autosampler:
•
Connecting the Solvent Line to the Accela Autosampler Heat Exchanger
•
Connecting the Solvent Line to the Accela Open Autosampler Injection Valve
Connecting the Solvent Line to the Accela Autosampler Heat Exchanger
The metal plate in front of the Accela Autosampler heat exchanger can reach temperatures as
high as 95
C (203
F).
Item
Third-party
part number
Thermo Fisher Scientific
part number
Ferrule, stainless steel, 0.19 in.
length, for 1/16 in. OD tubing
VICI Valco
ZF-1
00101-18122
Nut, 10-32 thread, 0.45 in.
length, for 1/16 in. OD tubing
IDEX
U-320
2522-1880
0.45 in.
0.19 in.
Ferrule
Nut
1/16 in. OD
stainless steel
tubing