Accela 600 Pump or Accela 1250 Pump Installation
Solvent Line Connections
Thermo Scientific
Accela 600 and 1250 Pump Hardware Manual
27
d. Reconnect the 20 μm particle size, stainless steel filter to the end of the solvent line
(see
Figure 13.
Solvent reservoir cap assembly with the solvent line assembly
e.
Screw the cap onto the solvent reservoir bottle until it is secure.
3. Position the bottles in the solvent bottle holder, allowing the solvent inlet lines to hang
down along the left side of the system:
• For an LC system with an Accela Autosampler, the solvent reservoir bottles reside in
the solvent platform at the top of the system.
• For an LC system with an Accela Open Autosampler, the solvent reservoir bottles
reside in the solvent tray attached to the autosampler table.
IMPORTANT
Use only stainless steel, 20 μm particle size sinker frits to
terminate the solvent reservoir lines.
Tip
The cap is a two-piece assembly. The upper section snaps onto a threaded
section. You can screw the threaded section onto the bottle and snap on the upper
section after installing the tube, or, if you are replacing existing tubing, you can
unscrew the entire cap from the bottle.
The ID of the upper section is 44 mm (1.75 in.). If necessary, you can use this
section alone for wide-mouth solvent bottles.
Solvent reservoir
bottle cap
Solvent reservoir
filter
Super Flangeless
fittings