Maintenance
Maintenance Schedule
78
Accela 600 and 1250 Pump Hardware Manual
Thermo Scientific
Maintenance Schedule
To ensure optimal performance of the pump, perform preventive maintenance at regularly
scheduled intervals. Perform corrective maintenance when you observe poor performance or
when you have been instructed to do so by a Thermo Fisher Scientific field service
representative.
lists the recommended maintenance schedule for the pump beginning with the most
frequently performed procedures.
Tip
With time, acetonitrile mobile phases coat the sapphire check valve seats with an
aliphatic amine residue that can cause the check valves to stick. To prevent residue
buildup, periodically pump HPLC-grade methanol or isopropanol (IPA) through the
system as described in
. If you are using a buffered
mobile phase, pump a non-buffered solvent that is miscible with the buffered mobile
phase through the system before pumping methanol or IPA through the system.
Table 12.
Maintenance frequency
Procedure
Frequency
Page
If the LC pump does not have the
optional seal wash pump, flush the LDA
once per day if you use buffered mobile
phases and once per month if you use
mobile phases that contain acetonitrile.
Flushing the Piston Guide Bushings
Once per day if you use buffered mobile
phases and you do not have the seal
wash pump installed
Replacing the Primary Piston Seals
Every 6 to 12 months depending on
usage
Every 6 to 12 months if you use buffered
mobile phases
Every 6 to 12 months depending on
usage
Cleaning the Instrument Casing
As needed
Removing Air from the Piston
Chambers
As needed
Cleaning the Displacement Chambers
and the Inlet Module
As needed
As needed
Replacing the Secondary Piston Seals
As needed