5
Maintenance
Accessing, Disassembling, and Reassembling the LDA
Thermo Scientific
Accela 600 and 1250 Pump Hardware Manual
89
Reassembling the LDA
Reassemble the LDA after replacing the piston seals or completing a corrective or
troubleshooting procedure that requires you to disassemble it.
To reassemble the LDA
1. Place the LDA on a clean, level surface.
2. Verify that the first check valve is in the appropriate orientation (see
), and then
insert it into the main body of the LDA (see
Figure 81.
Check valve orientation
Figure 82.
Inserting the displacement chambers, check valves, and outlet module into the LDA
CAUTION
The proportioning valve cables are hardwired to the proportioning valves.
Attempting to remove the cables from the valves can break the electrical wires inside
the insulated cables.
For rendering purposes only,
show the LDA without the cables.
Check
valve
Check
valve
Lower
displacement
chamber
Upper
displacement
chamber
Outlet
module