Reconnecting the Static Mixer and the High-Pressure Lines
To reconnect the static mixer and the high-pressure lines
1. Reinstall the static mixer as follows:
a. Using a 1/4 in. open-end wrench, reconnect the static mixer to the LDA.
b. Using a 1/4 in. open-end wrench, connect the static mixer to the inlet of the dynamic
mixer (see
Figure 35.
Static mixer connected to the LDA and the dynamic mixer
2. Connect the outlet of the dynamic mixer to the autosampler inlet (see
Solvent Line Between the Pump and the Autosampler”
).
LDA outlet
Static
mixer
Dynamic mixer
inlet