71
Program
Setting Display
Overview
Chain
Previous
Menu
LOAD
RS232
03/09/12
Timing Edit for Program
On-Time:
Run:
Memory
0.1
Skip
M001
Short-Time:
P/F-Time:
Off-Time:
Off
Off
Off
STEP:
01
Program
Off
Recall
Default
Program
number
Step number in
selected program
Program
settings
PROG:
01
Program
settings
Operation
1.
Press
FUNC
>
Program[F1].
Note that
Program[F1]
is off by default.
2.
Select
PROG
and select a program number to
edit.
PROG
01 ~ 16
3.
Select a
STEP
in the selected program.
STEP
01 ~ 16
4.
Select
Memory
and select which memory
location to load for the selected step.
Settings loaded from the memory location will
be used for the selected step.
The same memory location can be used for
multiple steps.
Memory
M001 ~ M256
5.
Set the
Run
setting for the step.
By default RUN is set to Skip.
The Auto setting will automatically start and go
onto the next step.
The Manual setting will wait for the user to
press
Next[F2]
before running the step.
Run
Skip, Auto, Manual
6.
Choose the
On-Time
in seconds.
The on-time setting determines how long the
load is turned on for the selected step.
The on-time is defined as the total test time
minus the off-time.
On-Time
0.1 ~ 60 seconds