148
Overpower protection(OPP)
0.1W ~ 192.5W
0.3W ~ 385W
1W ~ 1155W
or 110% of the maximum power of each range
Load off or limit selectable
Overheat protection(OTP)
Turns off the load when the heat sink temperature
reaches 95 °C
Undervoltage protection(UVP)
Turns off the load when detected. Can be set in the range
of 0 V to 150 V or Off.
Reverse connection protection(RVP)
By diode. Turns off the load when an alarm occurs.
Rating overcurrent protection (R.OCP)
An R.OCP message will be produced when the input
current range is greater than 110% of the rated operating
current range (I range).
Rating overpower protection (R.OPP)
An R.OPP message will be produced when the input
power range is greater than 110% of the rated operating
power range.
Front panel input rating overcurrent protection (F.R.OCP)
An F.R.OCP message will be produced when the front
panel input current range is greater than 77A (Typical).
7-7-13. Sequence
Normal Sequence
Operation mode
CC, CR, CV or CP
Maximum number of steps
1000
Step Execution Time
1ms
– 999 h 59 min
Time resolution
0.05ms (0.05ms
– 1min)
100ms (1min
– 1h)
1s (1h
– 10h)
10s (10h
– 100h)
1min (100h
– 999h 59min)
Fast Sequence
Operation mode
CC or CR
Maximum number of steps
1000
Step Execution Time
25 μs – 600 ms
Time resolution
1us(25us -60ms)
10us(60.01ms -600ms)