42
2-2-6. CC,CR and CP Mode Response Speed
Description
By default, the “normal current response” speed
is set to 1/1. The response speed can be
reduced to 1/2, 1/5, 1/10.
Reducing the current response speed can
affect other settings such as the slew rate and
soft start settings.
Operation
1.
Make sure the load is off.
2.
Press
Main
>
Configure[F5]
>
Other[F2]
and
set the
Response
parameter.
Range:
1/1,1/2,1/5,1/10
2-3. Advanced Configuration Settings
Use the advanced configuration settings to configure settings other
than those described in the basic configuration chapter.
2-3-1. Soft Start Setting
Description
The soft start setting is used to limit the amount
of input current at start-up or from when the Von
Voltage threshold is tripped.
The soft start setting only applies to CC & CR
mode.
Input I
Time
Soft Start = ON
Input I
Time
Soft Start= OFF
Rise voltage
Rise current
Rise current
Rise voltage
Operation
1.
Press
Main
>
Configure[F5]
>
Other[F2]
and
set the
Soft Start
time.
Range:
OFF, 1~200ms