49
Display
2.596
V
500.06
mA
1.2982
w
CV B Value
15.000
CV A Value
15.000
LOAD
RS232
A Value
Fine
03/09/12
V
V
OCP Alarm
Please Press Enter
To Clear Alarm
OCP
OCP indicator
Alarm message
when OCP is
set to Load Off
Mode
CV
I Range
L 0.7A
V Range
L 15V
Response
Fast
Configure
2-5-2. OPP
Description
For OPP, the LSG Series can be configured to
either limit the power or turn off the load.
The OPP levels can be set to 10% higher than
the rating power.
Operation
1.
Press
Main
>
Configure[F5]
>
Protection[F1]
and set the
OPP Level
and
OPP Setting
.
Range:
OPP Level: Rating power + 10%
OPP Setting: LIMIT, Load Off
Alarm
When
OPP Setting
is configured to
Load Off
, a
message will be displayed on the screen when
OPP is tripped. The Enter key must be pressed
to clear the alarm message.
When configured to
LIMIT
, OPP will be
displayed on the screen when the OPP is
tripped and the power will be limited to the
OPP
Level
setting.
Display
2.596
V
500.06
mA
1.2982
w
CV B Value
15.000
CV A Value
15.000
LOAD
RS 232
A Value
Fine
03/09/12
V
V
OPP Alarm
Please Press Enter
To Clear Alarm
Alarm message
when OPP is
set to Load Off
Mode
CV
I Range
L 0.7A
V Range
L 15V
Response
Fast
Configure
OPP
O
P
P indicator