44
Time
Von Latch = ON
Von
V
Time
I
V out
Load on
Operation
1.
Press
Main
>
Configure[F5]
>
Other[F2]
and
set the
Von Latch
setting.
Range:
Von Latch: OFF, ON
2-3-2-3. Von Voltage Delay
Description
Von Delay is the amount of time the unit will wait
before turning the load on after the Von Voltage
threshold has been latched. This will prevent
overshoot current from affecting the Von Voltage
threshold.
Operation
2.
Press
Main
>
Configure[F5]
>
Other[F2]
and
set the
Von Delay
time.
Range:
Von Delay: OFF, 0.5~60ms
CR mode can have the delay time set
separately
from the other modes (called Von Delay
–CR when
in CR mode).
2-3-3. Timer Functions
2-3-3-1. Count Time
Description
When Count Time is set to on, it will count the
elapsed time from when the load was turned on
to when it was turned off.
This function is applicable to manual and
automatic shutdown (such as from protection
functions such as UVP etc.)
The elapsed time will be shown in the display
Measurement area.
Operation
1.
Press
Main
>
Configure[F5]
>
Other[F2]
and
turn the
Count Time
on or off.
Range:
ON, OFF