108
CV Mode
Proportional Control:
Input voltage = rated voltage ×
(external resistance/10).
Inverse Control:
Input voltage = rated voltage ×
(1 - external resistance/10).
Input
Voltage
External
Resistance
Rated
Voltage
0
Ω
10k
Ω
Proportional
control
Inverse control
CP Mode
Proportional Control:
Input power = rated power ×
(external resistance/10).
Inverse Control:
Input power = rated power ×
(1 - external resistance/10).
Input
Power
External
Resistance
Rated
Power
0
Ω
10k
Ω
Proportional
control
Inverse control
Note
The inverse configuration is recommended for
safety reasons. In the event that any of the cables
become accidentaly disconnected, the
current/voltage/power input will drop to the
minimum. Under similar circumstances using
proportional control, an unexpectedly high input
would result.
Operation
1.
Turn the power off from the LSG Series and from
the load.
2.
Connect the external resistance across pins 1 and
3 of the J1 connector.
3.
Turn the power on the LSG Series.
4.
Set the operating mode and range.
5.
Press
Main
>
Configure [F5]
>
Next Menu [F4]
>
External [F3]
.
6.
Set the
Control
to
R
for proportional control or to
Rinv
for inverse control.