Remote Control
75
MEAS_SECX256
(0x08)
- x256 Secondary Gain
MEAS_SECX512
(0x09)
- x512 Secondary Gain
MEAS_SECX1024
(0x0A)
- x1024 Secondary Gain
MEAS_SECX2048
(0x0B)
- x2048 Secondary Gain
MEAS_SECX4096
(0x0C)
- x4096 Secondary Gain
Returns:
+OK:~: - indicates the gain was successfully set.
Comments:
If an invalid setting was specified, the configuration of the system is indeterminate.
MEAS_PRIAUTO and MEAS_SECAUTO are legacy constants and are not used by the system. If they are passed to
the function, they are ignored and the gain setting is not updated. If the function is called when auto ranging is
switched on, the autorange function is switched off and the gain is set to the current value.
Setup:SetAutORange
This sets the automatic ranging system on or off
Structure:
+S:O:<Autorange>:~:
Fields:
<Autorange> (integer) is the desired autorange state. If it is zero, auto ranging is
turned
off. If it is non-zero, auto ranging is turned on.
Returns:
+OK:~: indicates auto ranging state was updated.
Comments:
The autorange state is also affected by setting the measurement gain manually.
This automatically turns auto ranging off. The remote software should never assume the state of the auto ranging
system, but should explicitly set it whenever it is needed.
Setup:SetConfiguration
This sets the configuration of the test circuit.
Structure:
+S:C:<Configuration>:<Voltage>:~:
Fields:
<Configuration> (integer) is the configuration of the test circuit switching matrix. It is
the bit
wise combination of the specified HV and LV configurations as below:
MEAS_PRIH1-H3
- 0x00 - Energize H1 - H3
MEAS_PRIH2-H1
- 0x10 - Energies H2 - H1
MEAS_PRIH3-H2
- 0x20 - Energize H3 - H2
MEAS_PRH1-H2H3
- 0x30 - Energize H1 - H3 Link H2 - H3
MEAS_PRIH2-H1H3
- 0x40 - Energize H2 - H1 Link H1 - H3
MEAS_PRIH3-H1H2
- 0x50 - Energize H3 - H2 Link H1 - H2
MEAS_PRH1-H0
- 0x60 - Energize H1 - H0
MEAS_PRIH2-H0
- 0x70 - Energize H2 - H0
MEAS_PRIH3-H0
- 0x80 - Energize H3 - H0
MEAS_PRICAL
- 0x90 - Primary Calibration mode
MEAS_SECX1-X3
- 0x00 - Measure X1 - X3
MEAS_SECX2-X1
- 0x01 - Measure X2 - X1
MEAS_SECX3-X2
- 0x02 - Measure X3 - X2
MEAS_SECX1-X2X3
- 0x03 - Measure X1 - X3 Link X2 - X3
MEAS_SECX2-X1X3
- 0x04 - Measure X2 - X1 Link X1 - X3
MEAS_SECX3-X1X2
- 0x05 - Measure X3 - X2 Link X1 - X2
MEAS_SECX1-X0
- 0x06 - Measure X1 - X0
MEAS_SECX2-X0
- 0x07 - Measure X2 - X0
MEAS_SECX3-X0
- 0x08 - Measure X3 - X0
MEAS_SECCAL
- 0x09 - Primary Calibration mode
<Voltage>
(integer) is the required output voltage to energize the transformer.
This is one of:
Summary of Contents for TTR 2795
Page 1: ...Operating Instructions HAEFELY TEST AG TTR 2795 Transformer Turns Ratio Meter Version 1 4 ...
Page 10: ...6 Introduction Maximum Roll Diameter Outer 50mm Inner 10mm ...
Page 22: ...18 Connection and Setup Figure 11 Connecting to three phase BCT ...
Page 101: ...Application Software APSW 2795 97 ...
Page 111: ...Miscellaneous 107 14 7 Conformity ...