Remote Control
<NomTap> (integer) is the index of the nominal tap of the transformer.
<StepVal> (float) is the tap-to-tap step value, which is either in Volts or a
percentage of the nominal tap voltage, based on the Step Unit set. A negative step
value
indicates that the tap is on the HV side. A positive step value indicates that the tap
in on
the LV side. A step value of zero indicates that the tap voltages should be set
manually.
<MeasTap> (integer) indicates the number of taps that have actually been
measured. If a
test has been performed, and this parameter is 0, only a single winding has been
tested.
Comments:
None
Test:Results:Info
This returns the test information for the last test performed
Structure:
+T:R:I:~:
Fields:
None
Returns:
+OK:<Serial>:<Location>:<Type>:<Operator>:<Deviation>:<TimeDate>:~:
<Serial> (string) is the 20 character unit serial number
<Location> (string) is the 20 character unit location code
<Type> (string) is the 20 character manufacturers type code
<Operator> (string) is the 20 character operator code
<Deviation> (float) is the maximum acceptable turns ratio deviation.
<TimeDate> (float) is the Time and Date of the test.
Comments:
None
Test:Results:Taps
This returns the result for the individual taps on the transformer.
Structure:
+T:R:T:<TapNum>:~:
Fields:
<TapNum> (integer) is the number of the tap for which the results are to be shown.
If the
transformer is not tapped, this should be 0. The maximum value is <NumTaps>
Returns:
+OK:<HV>:<LV>:<TRA>:<IA>:<PA>:<TRB>:<IB>:<PB>:<TRC>:
<IC>:<PC>:<Pass>:~:
- The results of the test
+ERROR:0907:~:
- The value of <TapNum> is out of range
+ERROR:090E:~:
- The value of <NumTap> is valid, but that tap has not
been
measured. This will occur if the test was halted before completion
<HV> (float) is the boilerplate HV side voltage for this tap
<LV> (float) is the boilerplate LV side voltage for this tap
<TRA> (float) is the measured turns ratio for phase A
<IA> (float) is the measured energization current for phase A
<PA> (float) is the measured phase shift for phase A
<TRB> (float) is the measured turns ratio for phase B
Summary of Contents for TTR 2795
Page 1: ...Operating Instructions HAEFELY TEST AG TTR 2795 Transformer Turns Ratio Meter Version 1 4 ...
Page 10: ...6 Introduction Maximum Roll Diameter Outer 50mm Inner 10mm ...
Page 22: ...18 Connection and Setup Figure 11 Connecting to three phase BCT ...
Page 101: ...Application Software APSW 2795 97 ...
Page 111: ...Miscellaneous 107 14 7 Conformity ...