Remote Control
65
Comments
:
If the test system is idle, this will return the results of the last measurement made.
Once a test is started, but before any measurements are made, the system will report 0 for <TR>,<I> and <P>. If the
test was faulted, the function returns the appropriate fault code. This can be cleared by running another test or by
calling the Test:Measure:Halt command while the test system is idle.
Test:Measure:Continue
When testing a tapped transformer, this indicates the next tap is ready to be tested
Structure:
+T:M:C:~:
Fields:
None
Returns
+OK:~:
Comments
If the system is not waiting to test the next tap in sequence this command is
ignored.
Test:Results:Setup
Returns the basic setup parameters for the last test performed.
Structure:
+T:R:S:~:
Fields:
None
Returns
:
+OK:<VectorGroup>:<Voltage>:<HVNomVoltage>:
<LVNomVoltage>:<NumTaps>:<BotTap>:<NomTap>:
<StepVal>:<MeasTap>:~:
<VectorGroup> (integer) is the vector group of the transformer under test. This
defines
the configuration of the transformer and the phase displacement. Bits 15-12 indicate
the
HV winding configuration, bits 11-8 indicate the LV winding configuration and bits 7-
0
indicate the phase displacement. The possible values for the winding configurations
are:
0x00 – Delta
0x01 – Star, no neutral
0x02 – Star, With neutral
0x03 – Zig-zag, no neutral
0x04 – ZigZag, with neutral
0x05 – Single Phase (conventional transformer)
0x06 – Single Phase (current transformer)
0x0E – Use Range Extension transformer
The phase displacement is between 0 to 11.
<Voltage> (integer) indicates the voltage at which the transformer was tested.
It will be one of:
0x0A (10) - 10 Volts energization
0x28 (40) - 40 Volts energization
0x64 (100) - 100 Volts energization
<HVNomVoltage> (float) is the Voltage of the nominal tap on the HV Side.
<LVNomVolatge> (float) is the Voltage of the nominal tap on the LV Side.
<NumTaps> (integer) is the number of taps that have been defined for the
transformer.
<BotTap> (integer) is the index of the bottom tap of the transformer.
Summary of Contents for TTR 2795
Page 1: ...Operating Instructions HAEFELY TEST AG TTR 2795 Transformer Turns Ratio Meter Version 1 4 ...
Page 10: ...6 Introduction Maximum Roll Diameter Outer 50mm Inner 10mm ...
Page 22: ...18 Connection and Setup Figure 11 Connecting to three phase BCT ...
Page 101: ...Application Software APSW 2795 97 ...
Page 111: ...Miscellaneous 107 14 7 Conformity ...