Application Software APSW 2795
101
To continue with the arbitrary phase shift measurements, following information should be known.
(a) Type of the transformer to be tested.
-
Transformer having at least two three phase windings
-
Transformer having at least one winding in three phase and one winding in single-phase.
(b) Primary and secondary winding configuration of the transformer. For example, Delta-delta, Delta-star,
etc.
Example.
Transformer 3ph, Rated power 5500 kVA, HV: 11000+2x2.5 V; LV1: 2202V.
Vector group: Dd11.75 (or phase displacement=-7.5°)
Vector group: Dd 0.75 (or phase displa22.5°)
(Images courtesy: Trasfor, Lugano, Switzerland).
This is the case of a transformer having at least two three-phase windings, with primary and secondary side winding
of the transformer configured respectively as Delta and delta, the vector group settings on the primary and secondary
should be set-up as shown in the image below.