58
Remote Control
+ERROR:nnnn:~:
Data is encoded into the Fields: using a “big endian”, that is with the highest order bytes transmitted first. The Fields:
are encoded as follows:
Strings are transmitted with the terminating NULL character removed. The string should be scanned to determine the
position of control characters. If ‘:’, ‘~’, ‘+’ or ‘/’ occur within a string to be transmitted, they should be replaced with ‘/:’,
‘/~’, ‘/+’ or ‘//’ respectively. A string returned from the instrument should be scanned and if ‘/:’, ‘/~’, ‘/+’ or ‘//’ are
detected, they should be replaced with ‘:’, ‘~’, ‘+’ or ‘/’ respectively
Integers are passed as 4 (16 bit) or 8 (32bit) character ASCII coded hexadecimal strings. If an 8-byte string is coded
into 16-bit variable, only the most significant bytes are used.
Floating-point numbers are passed as an 8 character ASCII coded memory representation of a 4-byte IEEE format
float.
TimeDate information is passed as a 12 character string in a compressed format of YYMMDDHHMMSS.
11.3
Remote Control System Error Codes
The following error codes are defined for the TTR remote control system:
ES_MEMERR
(0x0901) - Error occurred processing memory request
ES_MEMUSED
(0x0902) - Requested memory is used already
ES_MEMFREE
(0x0903) - Requested memory has no data
ES_MEMCORRUPT
(0x0904) - Memory data is corrupted
ES_MEMOOR
(0x0905) - Memory Index out of range
ES_MEMFULL
(0x0906) - Memory is full
ES_TAPOOR
(0x0907) - Tap number out of range
ES_NOCONN
(0x0908) - Connection refused
ES_VECTORINV
(0x0909) - Vector group invalid
ES_VOLTINV
(0x090A) - Voltage Invalid
ES_BADBOTTAP
(0x090B) - Bottom tap invalid
ES_MEASRUN
(0x090C) - Measurement already running
ES_CANNOTRUN
(0x090D) - Measurement cannot run
ES_TAPNOTMEAS
(0x090E) - Tap has not been measured
ES_INVALIDREC
(0x090F) - Invalid Recipe index
ES_BADINDEX
(0x0910) - Bad Text Index
ES_CONFIGINV
(0x0911) - Switching Matrix Configuration Invalid
ES_CAL_IDXOOR
(0x0912) - Calibration factor index out of range
ES_CONFIGINV
(0x0913) – Configuration Invalid
ES_CALINVALID
(0x0914) - Calibration EEPROM checksum failed
ES_STEPPERINV
(0x0915) - Step Percent Value is an invalid value
ES_INVRECDATA
(0x0940) - Unrecognised data received.
ES_TESTRUNNING
(0x0300) - Test Procedure is running
These codes are used by the TTR to indicate the status of the system or any errors that have been detected.
11.4
Communications Functions
The communications functions are responsible for creating and managing connections with the unit. When a
communications link has been established, the remote system should ensure that the link is kept alive by sending a
Summary of Contents for TTR 2795
Page 1: ...Operating Instructions HAEFELY TEST AG TTR 2795 Transformer Turns Ratio Meter Version 1 4 ...
Page 10: ...6 Introduction Maximum Roll Diameter Outer 50mm Inner 10mm ...
Page 22: ...18 Connection and Setup Figure 11 Connecting to three phase BCT ...
Page 101: ...Application Software APSW 2795 97 ...
Page 111: ...Miscellaneous 107 14 7 Conformity ...